2310000381
LC COLUMN QUALITY CONTROL PERFORMANCE REPORT |
|
COLUMN INFORMATION SERIAL NUMBER : 2310000381 PART NUMBER : C0902-A3509 |
TEST CONDITIONS MOBILE PHASE : 70% Acetonitrile / 30% water COLUMN PRESSURE : 324Bar COLUMN FLOW : 0.80 ml / min LINEAR VELOCITY : 0.343 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 2.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 32858 MIN : 30000 SELECTIVITY : 1.82 RANGE : 1.78 - 1.88 USP TAILING FACTOR : 0.99 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.26 |
Peak Conc Sample (µg/ml) Component 1. 10 Uracil 2. 400 Phenol 3. 50 4-Chloro Nitrobenzene 4. 80 Toluene |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
ChraReport at https://chracert.com/2310000381 |