Chratec

Product Specification Sheet by ChraWork a Gimium™ Ecosystem
Skip to main content

2012000304

LC COLUMN
QUALITY CONTROL PERFORMANCE REPORT

SERIEAL NUMBER  : 2012000304

PART NUMBER       : C1505-B5032
COLUMN TYPE       : ChraSil B100 C18, 150 x 4.6 mm, 5.0um
LOT                         : VD184003

TEST CONDITIONS
MOBILE PHASE                   : 85% Acetonitril / 15% water
COLUMN PRESSURE           : 35Bar
COLUMN FLOW                  : 1.00 ml / min
LINEAR VELOCITY               : 0.205 cm / sec
TEMPERATURE                    : AMBIENT (Nominally 23 0C)
INJECTTION VOLUME         : 3.0 µl
TEST VALUES                                        SPECIFICATIONS
THEORETICAL PLATES : 17455              MIN      : 16780
SELECTIVITY                : 3.27                 RANGE  : 3.02 - 3.62
USP TAILING FACTOR  : 1.13                RANGE   : 0.98 - 1.25
(@ 5% Peak Height)                             k’           : .89
Peak                   Conc Sample (µg/ml)           Component
1.                                 5                                         Uracil
2.                                 200                                   Phenol
3.                                 25           4-Chloro Nitrobenzene
4.                                 850                                  Toluene

skd.jpg

MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST