LC Column QUALITY CONTROL PERFORMANCE REPORT |
|
SERIEAL NUMBER: 2012000304 PART NUMBER: C1505-B5032 |
TEST CONDITIONS MOBILE PHASE = 85% Acetonitril / 15% water COLUMN PRESSURE = 35Bar COLUMN FLOW = 1.00 ml / min LINEAR VELOCITY = 0.205 cm / sec TEMPERATURE = AMBIENT (Nominally 23 0C) INJECTTION VOLUME = 3.0 µl |
TEST VALUES _ SPECIFICATIONS THEORETICAL PLATES = 17455 MIN = 16780 SELECTIVITY = 3.27 RANGE = 3.02 - 3.62 USP TAILING FACTOR = 1.13 RANGE = 0.98 – 1.25 (@ 5% Peak Height) k’ = 2.89 |
Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |