# 2012000304
**LC COLUMN** **QUALITY CONTROL PERFORMANCE REPORT** | |
**COLUMN INFORMATION** SERIAL NUMBER : **2012000304** PART NUMBER : C1505-B5032 COLUMN TYPE : ChraSil B100 C18, 150 x 4.6 mm, 5.0um LOT : VD184003 | **TEST CONDITIONS** MOBILE PHASE : 85% Acetonitrile / 15% water COLUMN PRESSURE : 35Bar COLUMN FLOW : 1.00 ml / min LINEAR VELOCITY : 0.205 cm / sec TEMPERATURE : AMBIENT (normally 23 0C) INJECTTION VOLUME : 3.0 µl |
TEST VALUES SPECIFICATIONS THEORETICAL PLATES : 17455 MIN : 16780 SELECTIVITY : 3.27 RANGE : 3.02 - 3.62 USP TAILING FACTOR : 1.13 RANGE : 0.98 - 1.25 (@ 5% Peak Height) k’ : 2.89 | Peak Conc Sample (µg/ml) Component 1. 5 Uracil 2. 200 Phenol 3. 25 4-Chloro Nitrobenzene 4. 850 Toluene |
[](https://chratec.info/uploads/images/gallery/2021-08/skd.jpg) | |
MATERIAL SAFETY DATA SHEETS ARE AVAILABLE UPON REQUEST | |
[ChraReport](https://chrareport.com/2012000304 "2012000304") at https://chracert.com/2012000304 |